K 90 macro: two separate, finely serrated gold-plated contacts for current and voltage

Guide

Kelvin four-wire measurement: basics and choosing a clip

Measuring resistance in the milliohm range takes four wires, not two. Here is how Kelvin four-wire measurement works, and how to find the clip that suits your test object.

01 At a glance

The short answer

Kelvin or four-wire measuring leads are the contacting means you need to measure the resistance of contacts, coils and joints in the milliohm range with an LCR bridge. Each Kelvin clip brings two conductors to the test object: one for the measuring current, one for sensing the voltage.

To choose a clip: K 55 for the tightest spaces and the finest wires, K 90 as the standard for contacts, coils and joints, K 181 for power cables, busbars and battery terminals.

K 55 Kelvin clips contacting the thin wire of a toroidal coil

02 Principle

Four wires, not two

In a two-wire measurement, the measuring current flows through the same leads and contacts that are used to measure the voltage. Lead and contact resistances end up in the result, and in the milliohm range they carry real weight.

Four-terminal measurement separates the two. One pair of conductors drives the measuring current through the test object, the second pair senses the voltage right at the test object. Practically no current flows in the sense leads, so lead and contact resistances stay out of the reading.

On Schnepp Kelvin clips the conductors are labelled on yellow marking sleeves: +I and −I for the current, +U and −U for the voltage. The soft-PVC insulating sleeves come in red and black.

Current path +I / −I, Sense path +U / −U Current source V Voltage sense +I −I +U −U R Test object Kelvin clip Kelvin clip
Current path +I and −I drive the measuring current through the test object.
Sense path +U and −U pick up the voltage right at the test object.

03 Choosing a clip

The right Kelvin clip

Three sizes for three jobs. What matters is the space available, the cross-section of the test object and how delicate it is.
K 55, K 90 and K 181 Kelvin clips side by side to scale
The right Kelvin clip
Property K 55 K 90 K 181
Type small Kelvin clipstandard Kelvin cliplarge Kelvin clip
Use tightest spaces and housings, the finest coil and lead wires, low weight spares the test objectcontacts, coils and joints in the milliohm rangelarge transformers, power cables, conductor ropes, busbars, battery internal resistance on terminals
Length 55 mm90 mm181 mm
Jaw opening 5 mm8 mm40 mm, for cross-sections of 800 to 1000 mm²
Weight 7 g13 g122 g
Contact brass profile, finely toothed, hard-gold-plated, 6 × 2 mm, tips 2 × 2 mmbrass, finely toothed, gold-plated, 4 × 2 mm, tips 1 × 0.5 mmadaptive, form-fitting contact shape, grips large cross-sections without slipping
Contact force from insulating elastomer rings, ozone- and UV-resistantfrom insulating elastomer rings, ozone-resistantover 100 N, robust, slip-proof clip
Cable 1.5 m, coax RG 174 U or LiFY 0.5 mm²1.5 m, up to 10 m on request, coax RG 174 U or PVC strand 0.5 or 1.5 mm²3 m, PVC strand 1.5 mm² or silicone 1 mm², measuring current up to 20 A
Product page K 55K 90K 181
  • K 55

    Type
    small Kelvin clip
    Use
    tightest spaces and housings, the finest coil and lead wires, low weight spares the test object
    Length
    55 mm
    Jaw opening
    5 mm
    Weight
    7 g
    Contact
    brass profile, finely toothed, hard-gold-plated, 6 × 2 mm, tips 2 × 2 mm
    Contact force
    from insulating elastomer rings, ozone- and UV-resistant
    Cable
    1.5 m, coax RG 174 U or LiFY 0.5 mm²
    Product page
    K 55
  • K 90

    Type
    standard Kelvin clip
    Use
    contacts, coils and joints in the milliohm range
    Length
    90 mm
    Jaw opening
    8 mm
    Weight
    13 g
    Contact
    brass, finely toothed, gold-plated, 4 × 2 mm, tips 1 × 0.5 mm
    Contact force
    from insulating elastomer rings, ozone-resistant
    Cable
    1.5 m, up to 10 m on request, coax RG 174 U or PVC strand 0.5 or 1.5 mm²
    Product page
    K 90
  • K 181

    Type
    large Kelvin clip
    Use
    large transformers, power cables, conductor ropes, busbars, battery internal resistance on terminals
    Length
    181 mm
    Jaw opening
    40 mm, for cross-sections of 800 to 1000 mm²
    Weight
    122 g
    Contact
    adaptive, form-fitting contact shape, grips large cross-sections without slipping
    Contact force
    over 100 N, robust, slip-proof clip
    Cable
    3 m, PVC strand 1.5 mm² or silicone 1 mm², measuring current up to 20 A
    Product page
    K 181

Data from Schnepp product documents, as of 2026.

04 Point contacts

Kelvin probes for point contacts

Where no clip can grip, spring-loaded probe tips make the four-wire contact.
  • KP 100 Kelvin probe with slim black handle, collar and spring-loaded contact tip

    KP 100 Kelvin probe

    Four-wire lead with two spring-loaded probe tips for PCB tracks, IC pins, relays, potentiometers and precision-mechanical contacts. Contact heads KP 101, KP 103 and KP 104.

  • KP 200 Kelvin probe with black handle, red tip sleeve and gold-plated contact

    Custom: KP 200

    The KP 200 Kelvin hand probe is a customer-specific product of the Kelvin range: a robust hand probe for motors, transformers, busbars and oxidised contacts. An example of how we extend the range for a customer.

05 Leads and connectors

From test object to instrument

Compatible with LCR bridges and milliohmmeters. We fit the connector your instrument needs.
  • Shielded or highly flexible

    Shielded 50 Ω coax RG 174 U for fine measurements, or highly flexible PVC strand or silicone cable.

  • Twisted and braided

    The single cores are twisted together and held in a black polyamide braided sleeve.

  • Clearly marked

    Soft PVC insulating sleeves in red and black, yellow marking sleeves printed +U, −U, +I and −I, or coloured sleeves without print.

  • Your connector

    Gold-plated 4 mm plugs, BNC plugs or circular connectors.

  • EMC version

    K 55 and K 90 are insulated with PVC and contain no ferromagnetic parts.

  • Strain-relieved

    Strain relief and kink protection on every clip.

06 Special versions

Your measuring lead, your version

Like our standard products, Kelvin clip leads are also available to your specification. The K 90, for example, with a longer cable, an extra ground lead or the circular connector your instrument uses.

Small K 55 Kelvin clips with braided sleeving and four BNC plugs with yellow marking sleeves
  • Cables up to 10 m long
  • Additional ground lead
  • Connectors to match your instrument
  • Marking to your system

FAQ

Frequently asked questions

Which Kelvin clip suits battery internal resistance testing?

The large K 181 Kelvin clip. It is designed for measuring the internal resistance of batteries on their terminals, opens to 40 mm and grips with more than 100 N.

What do +U, −U, +I and −I mean?

They mark the conductors: +I and −I carry the measuring current, +U and −U sense the voltage. The marks are printed on the yellow marking sleeves of the Kelvin leads.

Which cable should I use for fine measurements?

Shielded 50 Ω coax RG 174 U. It is the standard cable for K 55 and K 90, with highly flexible strand as the alternative.

What measuring current is possible?

With the K 181 and its 1.5 mm² strand, a Kelvin measuring current of up to 20 A.

How long can the measuring lead be?

As standard, 1.5 m on the K 55 and K 90 and 3 m on the K 181. We make the K 90 with up to 10 m of cable as a special version.

Kelvin clips

The right clip for your test object

Tell us the clip, cable, length and connector, or describe your test object. We will come back with a quote.